EVALUATION OF SET-UP ERRORS IN 4D-DIBH THORACIC RADIOTHERAPY USING WHOLE-BODY VACLOCK AT VINMEC TIMES CITY INTERNATIONAL GENERAL HOSPITAL

https://doi.org/10.70755/vnjo.2025.78.10

Tóm tắt

.

Abstract

Purpose: To determine the daily initial set-up errors and the patient position reproducibility pre-irradiation with 4D-DIBH breathing control radiotherapy technique the whole-body Vac-lock immobilisation device.

Materials and methods: A total of 53 patients were assigned to the Thoracic Radiotherapy (Esophageal cancer: 31 patients; lung cancer: 15 patients; thymic tumor: 5 patients; Rib metastasis: 1 patient) at Vinmec Times City International Hospital from June 2023 to September 2024. Patients were treated by VMAT technique with 4D – DIBH (Deep Inhale Breath-Hold) using Whole Body Vac-lock immobilization devices (QFIX, PA, USA) and (CIVCO, Orange City, IA, USA). Determined the set – up error and pre-treatment repoducibility by collecting daily kV and CBCT images before treatment. A total 1027 of 3D-CBCT image serries were collected and analyzed offline on the ARIA system (Varian Medical System, CA, USA). These data were used to determined the setup margin (SM), systematic margin, and random margin errors in three directions: anterior-posterior (AP); left-right (LR); Cranial-Caudal (CC), and stored on the ARIA system (Varian Medical System, CA, USA).

Results: Innitial set – up error of 4D-DIBH radiotherapy for the thoracic tumors were: 2.4mm  ± 0.9mm (in the AP direction), 3.2mm  ± 1.3mm (in the Cranial-Caudal direction), and 1.9mm ± 1.1mm (in the LR direction). Average set-up error for the 4D-DIBH radiotherapy group of patients with esophageal cancer: 2.4 mm ± 0.9mm, 3.4 mm ± 1.2mm, and 1.8mm  ± 1.1mm in the AP, CC, and LR directions, respectively; average setting error for the 4D-DIBH radiotherapy group of patients with lung cancer: 2.7 mm ± 0.9mm (AP direction), 2.9 mm ± 1.3mm (CC direction), and 1.8mm  ± 0.9mm (LR direction). Average setting error for the 4D-DIBH radiotherapy group of patients with thymic cancer and ribs metastases: 1.8 mm ± 0.6mm, 3.0 mm ± 2.3mm, 3.1mm ± 1.4mm in the AP, CC, LR directions respectively; Set – up error (SM) for 4D-DIBH radiotherapy using Whole – Body Vac-lock is calculated according to the van Herk formula for the Thoracic region: 3.7mm (AP) 4.8mm (CC); 4.0mm (LR); esophageal cancer: 3.9mm (AP) 4.8mm (CC); 3.6mm (LR) ); lung cancer: 3.5mm (AP) 4.8mm (CC); 4.1mm (LR) ); thymic cancer, bone metastasis: 2.9mm (AP) 4.7mm (CC); 4.9mm (LR).

Conclusion: 4D-DIBH radiotherapy for the thoracic cancer using Body Vac-lock ensures well position repositioning, reduces initial set – up error to deliver accurate radiation dose to the tumors, and limits side effects on organ at risk (OAR). The results of the study and set – up error (SM) calculation is the base - evidence to help Radiation oncolosist determine the appropriate PTV - Margin for treatment in each location.

Từ khóa

.

Keywords

Setup error; 2D-kV image verification; cone-beam computed tomography (CBCT); immobilization systems; esophageal cancer; lung cancer; thoracic tumours; Deep Inspirarion Breath Hold (DIBH).

Tài liệu tham khảo (References)

  1. ICRU 62: International CommisCCon on Radiation Units and Measurements report 62. 
  2. van Herk M, Remeijer P, Rasch C, Lebesque JV (2000): The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys. 2000; 47(4):1121–35. 
  3. van Herk M (2004): Errors and margins in radiotherapy. Semin Radiat Oncol 2004; 14:52-64. 
  4. Hawkins M.A., Aitken A., Hansen V.N., McNair H.A., Tait D.M. (2011): Set-up errors in radiotherapy for oesophageal cancers – is electronic portal imaging or conebeam more accurate? Radiother Oncol J Eur Soc Ther Radiol Oncol. 2011; 98:249–254. 
  5. Yamashita H., Haga A., Hayakawa Y. (2010): Patient setup error and day-to-day esophageal motion error analyzed by cone-beam computed tomography in radiation therapy. Acta Oncol. 2010; 49:485–490. 
  6. Hashimoto T, Shirato H, Kato M, Yamazaki K, Kurauchi N, Morikawa T (2005):  Real-time monitoring of a digestive tract marker to reduce adverse effects of moving organs at risk (OAR) in radiotherapy for thoracic and abdominal tumors. Int J Radiat Oncol Biol Phys 2005; 61:1559–64. 
  7. Cohen R.J., Paskalev K., Litwin S., Price R., Feigenberg S.J., Konski A. (2010): Esophageal motion during radiotherapy: quantification and margin implications: esophageal motion during radiotherapy. Dis Esophagus. 2010; 23:473–479. 
  8. SaCCdharan S., Allison R., Jenkins T., Wolfe M., Mota H., CCbata C. (2005): Interfraction esophagus motion study in Image Guided Radiation Therapy (IGRT) Int J Radiat Oncol Biol Phys. 2005;63: S91–S92.  
  9. Gao X.-S., Qiao X., Wu F. (2007): Pathological analyCCs of clinical target volume margin for radiotherapy in patients with esophageal and gastroesophageal junction carcinoma. Int J Radiat Oncol Biol Phys. 2007;67: 389–396. 
  10. Dieleman E.M.T., Senan S., Vincent A., Lagerwaard F.J., Slotman B.J., van Sörnsen de Koste J.R. (2007) Four-dimenCConal computed tomographic analyCCs of esophageal mobility during normal respiration. Int J Radiat Oncol Biol Phys. 2007; 67:775–780.  
  11. Welch RW, Gray JE. (1982): Influence of respiration on recordings of lower esophageal sphincter pressure in humans. Gastroenterology. 1982; 83:590–594. 
  12. Good E, Oral H, Lemola K, et al. (2005): Movement of the esophagus during left atrial catheter ablation for atrial fibrillation. J Am Coll Cardiol. 2005; 46:2107–2110. [PubMed]. 
  13. Liliana Martins, Jose Guilherme Couto, Barbara Barbosa (2016): Use of planar kV vs. CBCT in evaluation of setup errors in oesophagus carcinoma radiotherapy. Oncol Radiother. 2016; 21(1): 57–62. 